Multicare Glucose Tests, 50 strips, Biochemical Systems
Quantitative whole blood glucose test.
For self-testing.
- SAMPLE: Capillary blood. Minimum volume: 0.5 mcL
- HEMATOCRIT RANGE: 30-55%
- MEASURING RANGE: 10-600 mg/dL (0.6-33.3 mmol/L)
- SENSITIVITY: Minimum measurable value: 10 mg/dL
- PERFORMANCE CHARACTERISTICS
- REPEATABILITY EVALUATION.
The multiCare-in glucose strips have demonstrated good performance.
Based on the results, the CV is:
- for glucose concentrations < 100 mg/dL, 2.4% to 4.3%
- for glucose concentrations ~ 100 mg/dL, 2.0% to 2.8%.
REPRODUCIBILITY ASSESSMENT (INTERMEDIATE PRECISION)
The multiCare-in glucose strips showed good performance in all ranges.
Three lots were evaluated, with the following CVs:
- Lot. 1: CV between 2.53% and 4.72%
- Lot. 2: CV between 2.66% and 4.29%
- Lot. 3: CV between 3.03% and 4.29%.
ACCURACY:
Comparison of the multiCare-in blood glucose monitoring system with the reference method using capillary whole blood samples collected simultaneously from 100 subjects.
1.1 Acceptability Criteria
Minimum acceptable accuracy requirements according to ISO 15197:2015
- 95% of individual blood glucose results must fall within ± 15 mg/dL (0.83 mmol/L) for glucose concentrations < 100 mg/dL (5.55 mmol/L)
- 95% of individual blood glucose results must fall within ± 15% for glucose concentrations ≥ 100 mg/dL (5.55 mmol/L).
1.2 Results
System accuracy results for glucose concentrations < 100 mg/dL (5.55 mmol/L)
- within ± 5 mg/dL (0.28 mmol/L)
- within ± 10 mg/dL (0.56 mmol/L)
- within ± 15 mg/dL (0.83 mmol/L)
- 80/150 (53.3%)
- 125/150 (86.6%)
- 145/150 (96.6%).
System accuracy results for glucose concentrations ≥ 100 mg/dL (5.55 mmol/L)
- within ± 5%
- within ± 10%
- within ± 15%
- 274/450 (60.8%)
- 411/450 (91.3%) 445/450 (98.8%).
System accuracy results for glucose concentrations between 28 mg/dL (1.5 mmol/L) and 580 mg/dL (32.2 mmol/L)
- within ± 15 mg/dL (0.83 mmol/L) or within ± 15% = 590/600 (98.3%) READING TIME: 5 seconds
STRIP TEMPERATURE: 10-40°C.
Ingredients:
Glucose Oxidase: 21 mcg; Mediator (Hexaammineruthenium chloride): 139 mcg; Buffer: 5.7 mcg; Stabilizer: 86 mcg.
Warnings:
For use with the meter.
Storage:
- Store at a temperature between 4° and 30°C
- EXPIRY AFTER OPENING: 90 days
- EXPIRY: 24 months.
Presentation: 50 strips
Features
| Product code | 69062 |
| Category | Blood glucose tests, Medical supplies, Useful kits, Focused Ads (IT) |
| Brand | Biochemical Systems |
| Quantity | 50 |
| Delivery from | Romania |
Multicare Glucose Tests, 50 strips, Biochemical Systems
-
Autoship & Save
- Enjoy 3% discount on your order
- Repeat deliveries can save you up to € 6,96 / year
- Skip, change or cancel anytime¹
Multicare Glucose Tests, 50 strips, Biochemical Systems€ 19,20 € 18,62Save € 0,58 with Autoship & SaveQuantity: -
Pay securely with your preferred payment method
Reviews
All reviews
All reviews
Alles bestens vielen dank
Read originalGood product for quality/price.
Right shipment on time .
Full OK!!! Full OK!!! Full OK!!!
Read originalSunteti f prompti si seriosi!Multumesc
Read original